Evaluation of SOFC single cell
 

 

Overview

BEL-SOFC is the automatic evaluation and deterioration test system for single plate cell in SOFC. Unique sample setting mechanism enables highly reproducible measurement.
Gas composition, gas flow and temperature can be controlled by dedicated software. Combining with electrochemical instruments, impedance, OCP and I-V properties can be measured.
Standard cell (Ni-YSZ |YSZ |LSM dia.:approx. 20mm) is prepared. It is useful to investigate the various performances and to compare with customer's cell.

 

Features

Evaluation and deterioration test for single cell. Full automatic system by dedicated software. Support software reduces the possibility of error in operation.

 

Specifications

Sample Single flat plate(dia.:approx. 20mm)
*Other diameters are negotiable Gases Anode Mixture of H
2, N2,CH4M,CO and saturated water vapor
Cathode Mixture of O
2 and N2
Gas flow control Mass flow controller x 2 (F.S. : 200sccm, Accuracy +/- 2%) Humidity Control Bubbling method(5-80
oC) Temperature Control Electric furnace:400-1000oC(accuracy: 0.5oC)
Thermostatic air oven:40-100
oC(accuracy: 0.5oC)
Bubbling + BEL-thermo:5-80
oC(accuracy: 0.2oC)
Flexible heater tube:50-100
oC(accuracy: 0.5oC) Pressure 100kPa (Abs) Electrochemical evaluation method Impedance measurement and I-V measurement by 3 terminal (Each electrode case) and 4 terminal pole (overall cell case) Condenser Set 15oC by circulation water bath Safety measure Short circuit breaker, lowering water level alarm, hydrogen detection alarm, CO detection alarm, Methane detection alarm Temperature highest limit alarm(heater OFF), Heater disconnection alarm (only for TIC1) Measurement software Condition setting:Measurement sequence setting (Temp.・flow・time),
Synchronism between the sequence setting and impedance measurement and I-V measurement.
Saving format:CSV file
Display:Flow chart/TIC1-4, Trend graph of MFC input value/I-VP result Physical W1100xD600xH1800mm(body:W750xD600xH1425mm)150 kg

 